Z-0783-2024: EliA GBM Wells, Material Number 14551401, for Phadia 250, 2500, and 5000. In vitro diagnostic

Recall record Z-0783-2024 for Phadia Ab. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0783-2024
Company in FDA record
Phadia Ab
FDA classification date
Jan 24, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Complaints that specific samples produced false positive EliA GBM results. A positive signal was present when these samples were tested for coating-solution reactivity using EliA wells without antigen.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter was sent to customers on 5/24/23. 1) Review instrument logfiles or LIS for EliA GBM results that fall in the range >10 EliA U/ml according to your internal procedures. Log files may only be available for analysis for a limited timeframe of the Phadia Laboratory System due to storage and maintenance restrictions and may not cover the entire timeframe of the Instrument message log. . Should you need help, contact Phadia US Technical Support who can further assist in collecting log files and aid in identifying the potentially impacted test results. 2) Until a resolution has been implemented and confirmed by Phadia AB, the use of the EliA GBM Well may be continued as detailed in the user manual and the DfU with the following recommendations: . For EliA GBM positive test results (>10 EliA U/ml): . Confirm positive EliA GBM results (>10 EliA U/ml) using an alternative method. If you do not have access to an alternative GBM method, please consider sending testing to your preferred reference laboratory for confirmation of positive results. . EliA GBM results d10 U/mL are not impacted by this issue and therefore these values can be reported according to the Interpretation of Test Results section on the EliA GBM DfU Customer Acknowledgement Must be returned prior to release of your next order of EliA GBM Wells If you have any questions concerning this issue, please contact Phadia US Technical Support at 1-800-346-4364, option #2.
Initiated
2023-05-24
Root cause
Nonconforming Material/Component
Distribution
US Nationwide.
Recall number
Z-0783-2024
Recall status
Open, Classified
CDRH record ID
204264
FDA classified
2024-01-24
Recall event number
93403

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗