Skip to content

Public FDA record

Z-0784-05: Trep-Chek Syphilis Test Kit, 960 tests, Catalog No. TP-960, Distributed by DiaSorin, Inc. 1951 Northwest

Recall record Z-0784-05 for Diasorin Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0784-05
Company in FDA record
Diasorin Inc.
FDA classification date
Apr 28, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Package insert incorrectly states that specimen diluent is ready to use, however specimen diluent bottle in kit states that the diluent must be diluted to one liter.

Source facts

Action
Customers were contacted via telephone, informed of the incorrect package insert, asked to destroy any incorrect inserts and a correct insert was faxed to customers.
Initiated
2005-03-16
Root cause
Other
Terminated
2006-07-22
Distribution
MA, CA, TX, KS, NY, FL
Recall number
Z-0784-05
Recall status
Terminated
CDRH record ID
38380
FDA classified
2005-04-28
Recall event number
31710

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗