Skip to content

Public FDA record

Z-0785-2013: Solara 3G Custom Manual Wheelchair Product Usage: To provide mobility to a person restricted to a sitti

Recall record Z-0785-2013 for Invacare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0785-2013
Company in FDA record
Invacare Corporation
FDA classification date
Feb 4, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The potential exists for the wheel to rotate freely despite engagement of hub brake.

Source facts

Action
Invacare sent an URGENT: RECALL INFORMATION Letter dated September 10, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to 1) Complete the Provider Response Card indicating your receipt and understanding of these instructions. 2) Contact your customers to inform them of the recall. It is recommended that you provide them with the enclosed Customer Letter. Set up an appointment to correct the customer's wheelchair as soon as possible. 3) Please examine the wheelchair to confirm it has one of the serial numbers listed above. The serial number can be found on the front lower crossbar located on the right while sitting in the chair. 4) Using the enclosed retrofit kit, make the necessary correction. Installation instructions can be found on page 109 of the Solara service manual, which is enclosed. 5) Document the correction of the wheelchair on the enclosed tracking sheet. 6) Using the enclosed pre-paid UPS mailing label and original box, return the replaced parts to the address listed on the label. 7) Fax or email the Provider Response Form and tracking sheet to 330-329-3458 or recall@invacare.com. For questions please contact Invacare Customer Service at 800-333-6900.
Initiated
2012-09-10
Root cause
Process control
Terminated
2014-05-13
Distribution
US Nationwide Distribution including the states of CA & MT.
Recall number
Z-0785-2013
Recall status
Terminated
CDRH record ID
115268
FDA classified
2013-02-04
Recall event number
64044

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗