Z-0785-2019: RESERVOIR "Y" ADAPTER, (a) BT920 (b) BT926 Suction and Anticoagulation assemblies are indicated for use

Recall record Z-0785-2019 for Medtronic Perfusion Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0785-2019
Company in FDA record
Medtronic Perfusion Systems
FDA record creation date
Jan 29, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Medtronic discovered that certain Y-connectors included in the suction lines have exhibited tearing and cracking. These defects could cause the Y-connectors to leak when in use, which could result in potential blood loss and a breach in the circuit.

Source facts

Action
The firm, Medtronic, began notifying US consignees on 27-Nov-2018 via hand delivery to be followed up with a UPS 2-Day communication titled "URGENT MEDICAL DEVICE RECALL" to each listed consignee that does not confirm hand delivery. Communications OUS also began 27-Nov-2018 using locally approved methods. The letter described the product, problem and actions to be taken. Customers were requested to do the following: 1. Identify and quarantine all unused affected product as listed in your inventory. 2. Return all unused affected product in your inventory to Medtronic. Contact Medtronic Customer Service at 1-800-854-3570 to initiate a product return. Your local Medtronic Representative can assist you in the return and replacement of this product as necessary. 3. Complete the enclosed Customer Confirmation Certificate and email to RS.CFQFCA@medtronic.com. If you have any questions about this action, please contact your Medtronic representative or call Director, Cardiac and Vascular Group, at 763-526-2513 or email: carlos.j.alfonzo@medtronic.com.
Initiated
2018-11-27
Root cause
Process control
Distribution
Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, France, Germany, Japan, and Sweden.
Recall number
Z-0785-2019
Recall status
Open, Classified
CDRH record ID
169578
FDA record created
2019-01-29
Recall event number
81702

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗