Open original FDA record ↗Recall record
- FDA identifier
- Z-0785-2022
- Company in FDA record
- Philips Respironics, Inc.
- FDA record creation date
- Mar 18, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
In 2018, Philips Respironics added foam replacement to the preventive maintenance schedule for these devices after discovering the polyester-based polyurethane (PE-PUR) foam may degrade into particles. This 2018 correction has been superseded by the device removal and rework described in Philips Respironics 2021 recall action for the same devices.
Source facts
- Action
- In 2018, Philips Respironics added foam replacement to the preventive maintenance schedule for these devices after discovering the polyester-based polyurethane (PE-PUR) foam may degrade into particles. This 2018 correction has been superseded by the device removal and rework described in Philips Respironics 2021 recall action for the same devices. A Field Communication for Trilogy Preventive Maintenance Update was distributed to customers 6/12/18. PRODUCT(S) AFFECTED All Trilogy models with Inlet Air Path Assembly PURPOSE / SCOPE This Field Communication announces: " The addition of the replacement of the Inlet Air Path Assembly and Removable Air Path Foam during the 10000 blower hour/24 month (whichever comes first) PM and 17,500-blower hour PM interval. " Trilogy Ventilator Maintenance Record revision to include the replacement date for the Air Inlet Path Assembly and Removable Air Path Foam. ACTION REQUIRED At both the 10000 blower hour/24 month (whichever comes first) and 17,500-blower hour PM intervals, the Inlet Air Path Assembly and Removable Air Path Foam are to be replaced. PROCEDURE For Inlet Air Path Assembly replacement procedures, refer to the Trilogy 100, 200, 202 & O2 Service & Technical Information Manual P/N 1002735 Section 11.2.3. For Removable Air Path Foam replacement procedures, refer to the Trilogy 100, 200, 202 & O2 Service & Technical Information Manual P/N 1002735 Section 11.2.4. IF you have any questions contact: Customer Service and Technical Support at 1-800-345-6443 or 724-387-4000 or email technical support at respironics.service@philips.com.
- Initiated
- 2018-06-12
- Root cause
- Device Design
- Distribution
- Global Distribution. US Nationwide.
- Recall number
- Z-0785-2022
- Recall status
- Open, Classified
- CDRH record ID
- 191978
- FDA record created
- 2022-03-18
- Recall event number
- 89584