Z-0785-2022: Trilogy 100, Trilogy 200, Garbin Plus, Aeris, LifeVent Continuous Ventilators The device is intended to

Recall record Z-0785-2022 for Philips Respironics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0785-2022
Company in FDA record
Philips Respironics, Inc.
FDA record creation date
Mar 18, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

In 2018, Philips Respironics added foam replacement to the preventive maintenance schedule for these devices after discovering the polyester-based polyurethane (PE-PUR) foam may degrade into particles. This 2018 correction has been superseded by the device removal and rework described in Philips Respironics 2021 recall action for the same devices.

Source facts

Action
In 2018, Philips Respironics added foam replacement to the preventive maintenance schedule for these devices after discovering the polyester-based polyurethane (PE-PUR) foam may degrade into particles. This 2018 correction has been superseded by the device removal and rework described in Philips Respironics 2021 recall action for the same devices. A Field Communication for Trilogy Preventive Maintenance Update was distributed to customers 6/12/18. PRODUCT(S) AFFECTED All Trilogy models with Inlet Air Path Assembly PURPOSE / SCOPE This Field Communication announces: " The addition of the replacement of the Inlet Air Path Assembly and Removable Air Path Foam during the 10000 blower hour/24 month (whichever comes first) PM and 17,500-blower hour PM interval. " Trilogy Ventilator Maintenance Record revision to include the replacement date for the Air Inlet Path Assembly and Removable Air Path Foam. ACTION REQUIRED At both the 10000 blower hour/24 month (whichever comes first) and 17,500-blower hour PM intervals, the Inlet Air Path Assembly and Removable Air Path Foam are to be replaced. PROCEDURE For Inlet Air Path Assembly replacement procedures, refer to the Trilogy 100, 200, 202 & O2 Service & Technical Information Manual P/N 1002735 Section 11.2.3. For Removable Air Path Foam replacement procedures, refer to the Trilogy 100, 200, 202 & O2 Service & Technical Information Manual P/N 1002735 Section 11.2.4. IF you have any questions contact: Customer Service and Technical Support at 1-800-345-6443 or 724-387-4000 or email technical support at respironics.service@philips.com.
Initiated
2018-06-12
Root cause
Device Design
Distribution
Global Distribution. US Nationwide.
Recall number
Z-0785-2022
Recall status
Open, Classified
CDRH record ID
191978
FDA record created
2022-03-18
Recall event number
89584

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗