Z-0786-2014: SIMVIEW NT and SIMVIEW NT with IDI: October 01, 2012 Siemens ceased Simview distribution. Manufactured

Recall record Z-0786-2014 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0786-2014
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Jan 17, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In a very rare occurrence an intended movement of the SimView NT simulation system components could not be stopped with usual hand control or the collision detection functionality.

Source facts

Action
Siemens sent an Urgent Field Safety Notice to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. What preventive measures can the user take? To stop the movement in this case, push the emergency button / stop button at control console or push the in-room emergency button. The emergency button/ stop button functionality is described in the User Manual. There is no preventive measure to avoid this situation. This letter is intended to bring this potential safety issue to your awareness and to ensure that your personnel are trained about the use of the emergency button / stop button inside the simulation room and on the control console. Always monitor any movement of the simulator components. The Update Instruction containing the Urgent Field Safety Notice will be delivered by the Siemens Customer Service Group, as per their procedure Siemens is working on a solution to mitigate the described situation. You will be informed as soon as the update distributing this solution is available. Please include this Field Safety Notice in your SimView NT Simulator System Owner Manual chapter Safety Advisory Letters where it should remain. In the interests of safety, we ask that you perform the described preventive measures and inform all affected personnel immediately. The relevant National Competent Authority has been informed of this Customer Safety Advisory Notice. We regret any inconvenience that this may cause, and we thank you in advance for your understanding. For further questions please call (925) 602-8157.
Initiated
2013-12-11
Root cause
Device Design
Terminated
2014-03-31
Distribution
Class II Recall - Worldwide Distribution - US Distribution including the states of California, Louisiana, Massachusetts, Pennsylvania, New York, Kentucky and Michigan., and the countries of Bulgaria, Germany, Belarus, Russian Fed., P.R. China, India, Republic Korea, Germany, Costa Rica, Argentina, Portugal, Greece, Bulgaria, Vietnam, Hungary, Italy, Palestinian Ter, Hungary, Jordan, Egypt Poland, Greece, Curacao, St Eus, Malaysia, Egypt, Malaysia, South Africa Turkey, Thailand, Qatar and Iraq.
Recall number
Z-0786-2014
Recall status
Terminated
CDRH record ID
124791
FDA classified
2014-01-17
Recall event number
67153

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗