Skip to content

Public FDA record

Z-0787-2022: Sorin Centrifugal Pump System, CP5, Drive Unit (Catalogue Item #60-01-04, UDI: 04033817900948) used in an

Recall record Z-0787-2022 for LivaNova Deutschland GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0787-2022
Company in FDA record
LivaNova Deutschland GmbH
FDA record creation date
Mar 15, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

One ferrite ring expected to be installed on centrifugal pump system power cable may not be present.

Source facts

Action
On 12/17/2021, the firm notified affected customers via email using a letter titled, "URGENT MEDICAL DEVICE CORRECTION FA-CP-MUN-2021-001." The letter indicated the following actions to be taken by the customer: Using the attached Customer Response Form (Attachment 1), please respond by email to LivaNova.FSCA@livanova.com to confirm the receipt of this letter and that you have read and understood its content. No further action is requested. A LivaNova authorized technician will contact you to schedule an appointment and correct the devices with the missing ferrite ring. Any affected device currently in use may continue to be used without limitation. For questions regarding this Medical Device Correction, please contact a LivaNova sales rep or send an e-mail LivaNova Quality Assurance Team at LivaNova.FSCA@livanova.com. Please, complete this response form and return it via e-mail to LivaNova.FSCA@livanova.com no later than December 31st, 2021.
Initiated
2021-12-17
Root cause
Process control
Terminated
2024-01-08
Distribution
US distribution in states of FL, TX, and NE
Recall number
Z-0787-2022
Recall status
Terminated
CDRH record ID
191249
FDA record created
2022-03-15
Recall event number
89312

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗