Increased temperature in the probe-head surface with LOGIQ P6 Model BT07 and BT09 scanners when used in specific types of scans and with specific probes.
Source facts
Action
The firm disseminated its medical device correction notices by letter on 12/28/2018. The letter recommends discontinuing scanning patients in endocavity, surgical, and neonatal exams and to discontinue the use of E8C, E8CS, BE9C, BE9CS, 4DE7C, i12L, 8C, 4D8C, 7S, 5S, and 5Sp probes.
Initiated
2018-12-28
Root cause
Device Design
Terminated
2020-10-05
Distribution
Worldwide distribution.
Recall number
Z-0788-2019
Recall status
Terminated
CDRH record ID
170350
FDA classified
2019-01-30
Recall event number
81937