Open original FDA record ↗Recall record
- FDA identifier
- Z-0788-2022
- Company in FDA record
- Microbiologics Inc
- FDA record creation date
- Mar 16, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
LYFO Disk Catalog #0983L is to contain Candida tropicalis. However, lot number 983-40-3 under Catalog #0983L, labeled as Candida tropicalis, is actually Yersinia entercolitica. This labeling error may cause a user's quality control to fail, resulting in testing needing to be repeated and delayed test results.
Source facts
- Action
- Microbiologics notified customers on 07/17/2015 via letter titled "Medical Device Recall". Instructions included to inspect inventory for affected products. If customers have received the product identified within the recall (LYFO DISK 0983L, Lot# 983-40-3), it is recommended that the customer destroys the product or return it to the recalling firm. Acknowledgement of the receipt of the recall letter was requested by completing a form and returning to Microbiologics. Questions or concerns can be directed to the Quality Systems Manager at 320.229.7050.
- Initiated
- 2015-07-17
- Root cause
- Labeling mix-ups
- Distribution
- Distribution is nationwide with three consignees located in Colorado, Georgia, and Ohio.
- Recall number
- Z-0788-2022
- Recall status
- Completed
- CDRH record ID
- 191821
- FDA record created
- 2022-03-16
- Recall event number
- 89513