Skip to content

Public FDA record

Z-0788-2024: Power Knee, REF: PKA10001, and PKA10003; with Power Knee Battery, REF: PKA10002

Recall record Z-0788-2024 for Ossur H / F. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0788-2024
Company in FDA record
Ossur H / F
FDA classification date
Jan 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Knee batteries may need replacement with conforming batteries to ensure there isn't a failure to turn on the device, or an unexpected shutdown, which is a momentary loss of power to the knee where the user may experience loss of support that can lead to a fall and potential injury.

Source facts

Action
On 12/26/23, recall notices were mailed to customers who were asked to do the following: a. Check the serial number of the Power Knee Batteries to confirm the affected units. b. Replace the affected batteries with the new batteries supplied along with this letter. c. To ensure future traceability, please fit each new battery into the designated knee according to the knee Serial Number indicated on the sticker on the back of the battery. d. PASS THIS NOTICE to those who need to be aware within your organization. If you have further distributed this product, please identify your customers, and notify them at once of this product alert. We recommend that you include a copy of this notice. e. Complete and return acknowledgement forms and affected batteries in pre-paid return packaging provided. Customers with questions can contact customer service at 1-800-233-6263, extension 10
Initiated
2023-12-26
Root cause
Process control
Distribution
US Nationwide distribution including in the states of MA, SC, MD, TN, ND, VA, IL, NJ, FL, MI, NY, AL, TX, OK, IN, GA, CA, PA, NM, VT, NV, MS, HI, OH, WI, LA, IA, NE, MN, AR, OR, NC, NH, WV, CO, ID, AZ, MO, CT, WA, KY, UT, KS.
Recall number
Z-0788-2024
Recall status
Open, Classified
CDRH record ID
204706
FDA classified
2024-01-25
Recall event number
93549

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗