Open original FDA record ↗Recall record
- FDA identifier
- Z-0789-03
- Company in FDA record
- Medtronic Ave Inc
- FDA classification date
- May 7, 2003
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
During use, device may exhibit high insertional resistance which may result in device damage, kinking and subsequent balloon deflation difficulties.
Source facts
- Action
- A letter was sent to each US commercial accounts via UPS overnight mailing on Saturday, March 1st, 2003 for delivery on Monday, March 3rd 2003. The notification instructs firm''s customers to sequester product from the specific lot numbers so that firm''s rep could assist in removing the affected product.
- Initiated
- 2003-02-27
- Root cause
- Other
- Terminated
- 2003-09-22
- Distribution
- Product was released for distribution to 446 consignees nationwide. The recall was appropriately extended to the user level; i.e., Clinics, hospitals and medical Centers who received the recalled product. There were no shipments to U.S. Government facilities or foreign or Canadian distribution.
- Recall number
- Z-0789-03
- Recall status
- Terminated
- CDRH record ID
- 26485
- FDA classified
- 2003-05-07
- Recall event number
- 25794