Skip to content

Public FDA record

Z-0789-03: Medtronic AVE brand BridgeAssurant Balloon Expandable Stent Delivery System for Biliary Catheter;Model #s

Recall record Z-0789-03 for Medtronic Ave Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0789-03
Company in FDA record
Medtronic Ave Inc
FDA classification date
May 7, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

During use, device may exhibit high insertional resistance which may result in device damage, kinking and subsequent balloon deflation difficulties.

Source facts

Action
A letter was sent to each US commercial accounts via UPS overnight mailing on Saturday, March 1st, 2003 for delivery on Monday, March 3rd 2003. The notification instructs firm''s customers to sequester product from the specific lot numbers so that firm''s rep could assist in removing the affected product.
Initiated
2003-02-27
Root cause
Other
Terminated
2003-09-22
Distribution
Product was released for distribution to 446 consignees nationwide. The recall was appropriately extended to the user level; i.e., Clinics, hospitals and medical Centers who received the recalled product. There were no shipments to U.S. Government facilities or foreign or Canadian distribution.
Recall number
Z-0789-03
Recall status
Terminated
CDRH record ID
26485
FDA classified
2003-05-07
Recall event number
25794

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗