Abbott has identified an issue with the Alinity i Gear Pump Assembly (part number A 30108552 01) resulting in foaming/bubbling out of the bottle reservoir for Concentrated Wash Buffer and an unexpected amount of dried residue of buffer. The foaming/bubbling is a result of an improper calibration of the pump s operating speed.
Source facts
Action
The firm initiated the recall by letter on 12/20/2018. The letter identified the affected product, problem and actions to be taken. The units will be serviced in the field. For questions contact Customer Service at 1-877-4ABBOTT. Customers outside the U.S., contact your local area Customer Service.
Initiated
2018-12-20
Root cause
Process control
Terminated
2022-09-19
Distribution
Worldwide Distribution
Recall number
Z-0789-2019
Recall status
Terminated
CDRH record ID
170295
FDA record created
2019-01-30
Recall event number
81916