Open original FDA record ↗Recall record
- FDA identifier
- Z-0790-2014
- Company in FDA record
- Becton Dickinson & Co.
- FDA classification date
- Jan 17, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An incorrect reagent was packed with an in vitro diagnostic kit, which may cause an increase in unresolved results or false negative results for Clostridium difficile in patient samples.
Source facts
- Action
- Becton Dickinson sent an Urgent Product Recall Letter dated October 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. They were advised that a portion of the BD GeneOhm Cdiff assay test kits from one lot, 08T13108, may contain incorrect bags of sample buffer tube. Customers were requested to return response forms indicating acknowledgement of the recall and quantity of remaining product that was discarded. Please fax to 410-316-4258.
- Initiated
- 2013-10-11
- Root cause
- Error in labeling
- Terminated
- 2014-02-19
- Distribution
- US Distribution including the states of NY, IN, UT, MN, OH, MA, SD, CA, NJ, and PA.
- Recall number
- Z-0790-2014
- Recall status
- Terminated
- CDRH record ID
- 124243
- FDA classified
- 2014-01-17
- Recall event number
- 66933