Z-0790-2024: Bard Nasogastric Sump Tube with ENFit, 10Fr., 36in Long, REF EN0042100

Recall record Z-0790-2024 for C.R. Bard Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0790-2024
Company in FDA record
C.R. Bard Inc
FDA classification date
Jan 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

BD (C.R. Bard) has received user complaints reporting that during use of BD Bard NG Sump Tubes with ENFit Connectors, inadequate or reduced suction, decompression, and drainage occurred.

Source facts

Action
BD (C.R. Bard) issued an URGENT Medical Device Recall Notice to its consignees via FedEx and email. The notice explained the problem with the devices, potential risks, and the devices be removed, quarantined, and discarded. Disseminate the notice to all users within the facility. If you have purchased this product through a distributor, please contact this distributor to request appropriate credit or order alternate replacement product. Complete and return the attached Customer Response Form to the BD contact noted on the form confirming acknowledgement of the correction notification, whether or not you have any affected product, so that BD may confirm your receipt of this notification and provide credit or alternate product. In addition to the information above, Distributors were directed to take the following actions: Please discontinue distribution of all product codes in the table above, and immediately dispose of all affected product remaining in your possession in accordance with your local facility s process. 2. Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this field action notification on BD s behalf. Attachment 2 3. Complete the attached Distributor Response Form and return to the BD contact noted on the form confirming acknowledgement of the recall notification, whether or not you have any of the impacted product, so that BD may acknowledge your receipt of this notification. 4. Indicate on the response form the quantity of product identified at your facility and confirm that this inventory was destroyed. Attachment 1 5. Please contact your BD representative if you require assistance with credits to be issued for discarded products or for replacement product options. You may also contact the number below for further assistance.
Initiated
2023-12-20
Root cause
Under Investigation by firm
Distribution
US
Recall number
Z-0790-2024
Recall status
Open, Classified
CDRH record ID
205099
FDA classified
2024-01-25
Recall event number
93679

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗