The DoseRight feature suggest a mAs based on the measured patient size, a reference size and a reference mAs. When scanning large children, the suggested mAs may be higher than clinicians would expect.
Source facts
Action
Philips Medical Systems sent a Field Safety Notice letter dated February 25, 2011, to all affected customers.The letter identified the product the problem and the action needed to be taken by the customers.
Your Corrective Action Plan (CAP) is approved and consists of:
1. Notifying all affected users with Field Safety Notice 72800527/72800528 starting March 8, 2011. This Field Safety Notice has been reviewed and is approved.
2. Following up with all responses by dispatching a Field Service engineer to each site to install the software update at no cost to the user.
3. Tracking the responses from users by certified mail.
For further questions please call (440) 483-7600.
Initiated
2011-03-07
Root cause
Software design
Terminated
2014-04-28
Distribution
Worldwide Distribution - US Distribution including the states of DC, IL, IN, MI, MO, NY, OH, OR, PA, TX and VT., and the countries of Australia, Belgium, China, Germany, India, Israel, Italy and The Netherlands.
Recall number
Z-0791-2014
Recall status
Terminated
CDRH record ID
122030
FDA classified
2014-01-29
Recall event number
66311