Skip to content

Public FDA record

Z-0791-2019: 3M(TM) Surgical Clipper by Remington(R) REF 9602 The 3M Surgical Clipper Classic, 9602, is intended to r

Recall record Z-0791-2019 for 3M Company - Health Care Business. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0791-2019
Company in FDA record
3M Company - Health Care Business
FDA record creation date
Jan 30, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

3M has recently received reports involving the 3MTM Surgical Clipper by Remington, Model 9602, overheating in mobile/field EMS situations while charging from a modified sine wave power inverter.

Source facts

Action
The firm, 3M Health Care, disseminated their "Field Safety Action" letter dated 1/9/2019 to its customers beginning on 01/14/2019. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: -Consult facility bio-med or power inverter manual/manufacturer to ensure only pure sine wave inverters are being used in mobile/field EMS situations to charge the 9602 clipper, as emphasized in the new product caution. -Complete and return the Customer Acknowledgement Form to show your understanding of and compliance with this Field Safety Action via Email to: 3Mhealthcare@montage72.com. If you have questions regarding this matter, please contact your local 3M sales rep or the 3M Health Care Customer Helpline at 1-800-228-3957, option #6, Monday through Friday, 7:30 a.m. 6:00 p.m. Central Time.
Initiated
2019-01-14
Root cause
Nonconforming Material/Component
Terminated
2021-02-11
Distribution
US Distribution and countries Brazil and Canada.
Recall number
Z-0791-2019
Recall status
Terminated
CDRH record ID
170345
FDA record created
2019-01-30
Recall event number
81933

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗