Reason stated in the source
3M has recently received reports involving the 3MTM Surgical Clipper by Remington, Model 9602, overheating in mobile/field EMS situations while charging from a modified sine wave power inverter.
Source facts
- Action
- The firm, 3M Health Care, disseminated their "Field Safety Action" letter dated 1/9/2019 to its customers beginning on 01/14/2019. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: -Consult facility bio-med or power inverter manual/manufacturer to ensure only pure sine wave inverters are being used in mobile/field EMS situations to charge the 9602 clipper, as emphasized in the new product caution. -Complete and return the Customer Acknowledgement Form to show your understanding of and compliance with this Field Safety Action via Email to: 3Mhealthcare@montage72.com. If you have questions regarding this matter, please contact your local 3M sales rep or the 3M Health Care Customer Helpline at 1-800-228-3957, option #6, Monday through Friday, 7:30 a.m. 6:00 p.m. Central Time.
- Initiated
- 2019-01-14
- Root cause
- Nonconforming Material/Component
- Terminated
- 2021-02-11
- Distribution
- US Distribution and countries Brazil and Canada.
- Recall number