Skip to content

Public FDA record

Z-0792-05: Terumo Clirans Series EE Excebrane Hollow Fiber Dialyzers; a wet model single use dialyzer consisting of

Recall record Z-0792-05 for Asahi Medical Co Ltd. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0792-05
Company in FDA record
Asahi Medical Co Ltd
FDA classification date
May 3, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Exposure to freezing temperatures may damage the hollow fiber dialyzers and result in blood leaks during dialysis.

Source facts

Action
Asahi sent Henry Schein, Inc. a recall letter dated 3/30/05, requesting the recall of the affected lots of Asahi labeled product, and Henry Schein, Inc. recalled the Asahi labeled dialyzers by letter dated 3/31/05, with a copy of the Asahi 3/30/05 letter. The accounts were informed that the dialyzers may have been exposed to temperature variances during distribution, and werw requested to cease distribution/use of the specific lots listed and return them to Henry Schein for credit. The recall was expanded to include the Terumo labeled dialyzers distributed by Asahi by Henry Schein letter dated 4/13/05, accompanied by a Terumo letter dated 4/8/05, sent to the accounts via first class mail on 4/14/05. The instructions were the same as in the 3/31/05 letter. A revised letter dated 4/21/05 was sent on 4/25/05, correcting the Henry Schein product codes involved in the Terumo labeled product. Any questions were directed Asahi Kasei Medical Co. at 847-498-8500.
Initiated
2005-03-31
Root cause
Other
Terminated
2006-07-10
Distribution
Nationwide and Puerto Rico
Recall number
Z-0792-05
Recall status
Terminated
CDRH record ID
38487
FDA classified
2005-05-03
Recall event number
31536

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗