Open original FDA record ↗Recall record
- FDA identifier
- Z-0792-05
- Company in FDA record
- Asahi Medical Co Ltd
- FDA classification date
- May 3, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Exposure to freezing temperatures may damage the hollow fiber dialyzers and result in blood leaks during dialysis.
Source facts
- Action
- Asahi sent Henry Schein, Inc. a recall letter dated 3/30/05, requesting the recall of the affected lots of Asahi labeled product, and Henry Schein, Inc. recalled the Asahi labeled dialyzers by letter dated 3/31/05, with a copy of the Asahi 3/30/05 letter. The accounts were informed that the dialyzers may have been exposed to temperature variances during distribution, and werw requested to cease distribution/use of the specific lots listed and return them to Henry Schein for credit. The recall was expanded to include the Terumo labeled dialyzers distributed by Asahi by Henry Schein letter dated 4/13/05, accompanied by a Terumo letter dated 4/8/05, sent to the accounts via first class mail on 4/14/05. The instructions were the same as in the 3/31/05 letter. A revised letter dated 4/21/05 was sent on 4/25/05, correcting the Henry Schein product codes involved in the Terumo labeled product. Any questions were directed Asahi Kasei Medical Co. at 847-498-8500.
- Initiated
- 2005-03-31
- Root cause
- Other
- Terminated
- 2006-07-10
- Distribution
- Nationwide and Puerto Rico
- Recall number
- Z-0792-05
- Recall status
- Terminated
- CDRH record ID
- 38487
- FDA classified
- 2005-05-03
- Recall event number
- 31536