Table Bases may have been assembled incorrectly, missing a retaining ring from the head end assembly. Over time, the parts could go out of alignment, making clutch engagement or disengagement difficult. If the clutch cannot disengage, the user would be unable to rotate the patient from supine to prone.
Source facts
Action
All consignees were notified by telephone and a recall letter was mailed to each facility as a follow up. Customers were asked to remove tables from operation until they were inspected by OSI service personnel.
Initiated
2015-12-14
Root cause
Employee error
Terminated
2016-02-10
Distribution
US, South Africa, Japan and United Arab Emirates.
Recall number
Z-0792-2016
Recall status
Terminated
CDRH record ID
142535
FDA classified
2016-02-09
Recall event number
72937