Skip to content

Public FDA record

Z-0792-2020: Cochlear Baha 5 Sound Processor Tamper Proof Battery Door (3 pcs.)

Recall record Z-0792-2020 for Cochlear Americas Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0792-2020
Company in FDA record
Cochlear Americas Inc.
FDA record creation date
Jan 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product was packaged with the Incorrect tamper proof battery door.

Source facts

Action
On 12/18/2018, the firm sent a "MEDICAL DEVICE RECALL NOTICE" to affected customers via telephone or email informing them of a packaging error for a specific lot (LOT # COH1253733) Tamper Proof Battery Doors. Customers are instructed to: 1-Examine their inventory and if they should have any of the affected product in their possession to carefully verify that they are the tamper proof version, and if in doubt about the functionality or identity - to scrap the item and to inform the Recalling Firm by filling out and return the Response Form. The Recalling Firm will arrange to send replacement product. 2-If customers receive complaints or queries from their users about "loose battery doors" or similar, to please advise the user to check the functionality according to the Recalling Fir's instructions "Baha 5 User Manual" and to scrap and replace as necessary. 3-To identify customers or users to whom they have distributed battery doors from the affected lot and to provide contact information to the Recalling Firm. The Recalling Firm will make contact, explain the issue, and send replacement product directly to them if necessary. NOTE: If you (the customer) do not have specific records showing with customers may have received affected product, they are to consider providing the Recalling Firm with a list of all customers that received a Baha 5 system between Sept. 1, 2018 and Dec. 18, 2018 so that the Recalling Firm may contact them. For any questions about the product recall customers can contact the Recalling Firm's Quality Department at 303-264-2163.
Initiated
2018-12-18
Root cause
Packaging process control
Terminated
2021-01-28
Distribution
Domestic: CA, FL, IL, IN, MA, MO, NY, SD, TN, and TX OUS Regions; Asia Pacific (APAC), and Europe, The Middle East and Africa (EMEA)
Recall number
Z-0792-2020
Recall status
Terminated
CDRH record ID
176847
FDA record created
2020-01-24
Recall event number
83970

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗