Skip to content

Public FDA record

Z-0792-2025: Medline custom medical procedure kits labeled as: 1) ENSEMBLE RACHIS-LF, REF DYNJ47568B; 2) UROLOGY RO

Recall record Z-0792-2025 for MEDLINE INDUSTRIES, LP - Northfield. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0792-2025
Company in FDA record
MEDLINE INDUSTRIES, LP - Northfield
FDA classification date
Dec 30, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.

Source facts

Action
Medline issued a MEDICAL DEVICE VOLUNTARY RECALL notice on 11/21/2024 to its consignees via US Mail and email. The notice explained the issue, potential risk to health, and requested that the affected product be quarantined. Once the response is received, Medline will provide labels to place on affected inventory, with instructions for the consignee to remove the affected component prior to using the kit.
Initiated
2024-11-21
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution - US Nationwide and the countries of UAE, and Canada.
Recall number
Z-0792-2025
Recall status
Open, Classified
CDRH record ID
211386
FDA classified
2024-12-30
Recall event number
95852

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗