Z-0793-2019: GE Healthcare Centricity Universal Viewer Breast Imaging Centricity Universal Viewer is a device that

Recall record Z-0793-2019 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0793-2019
Company in FDA record
GE Healthcare, LLC
FDA record creation date
Feb 1, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When switching back & forth between multiple UV instances in the Windows taskbar, the patient images displayed on the mammo high resolution monitors may not show images of the patient selected on UV, which could result in an incorrect diagnosis.

Source facts

Action
On December 21, 2019 GE Healthcare sent "URGENT MEDICAL DEVICE CORRECTION" letters to all their customers stating the following: RE: Centricity Universal Viewer Breast Imaging workflow integration - Patient out of Context Please ensure that all potential users in your facility are made aware of this safety notification and the recommended actions. Safety Instructions You can continue to use your Centricity Universal Viewer. Follow the below safety instructions while performing your exams. Place the below instructions with your user manual until you receive the addendum. 1.Visually verify that the patient images and selection displayed on the UV monitor and the high-resolution review monitors are the same patient. Make sure that the patient name is displayed onall monitors to identify that the patient selection and images belong to the same patient. 2.When switching back & forth between multiple viewer instances, use the Windows taskbar orAlt+Tab to select a Universal Viewer instance. Click the SenoIris or IDI icon on the Windows taskbarto bring the SenoIris or IDI Mammo Workstation into focus. 3.To prevent this situation from occurring, always close the Universal Viewer instance before opening any new viewer instances. Affected Product Details Centricity Universal Viewer version 6.0 SP4 or higher Breast Imaging integrated with either IDI Mammo Workstation version 4.7.x or all versions of SenoIris. This issue does not affect Universal Viewer Native Breast Imaging. Universal Viewer GTIN 00840682103800. Product Correction GE Healthcare will provide an addendum to the Centricity Universal Viewer User Manual for your system. No other correction is required. Contact Information If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2018-12-20
Root cause
Software design
Terminated
2021-04-09
Distribution
Worldwide Distribution: US (nationwide) in states of: AZ, CA, CO, DE, FL, IA, IL, IN, KS, MA, ME, MI, MN, MO, MS, MT, NE, NJ, NV, NY, OH, PA, PR, SC, SD, TN, TX, VA, VT, and WI; and countries of: Antigua, Australia, Austria, Belgium, Bermuda, Brazil, Canada, Cayman Islands, Chile, Colombia, Cote d'Ivoire, France, Germany, Guatemala, India, Italy, Korea, Kuwait, Lebanon, Malaysia, Malta, Mexico, Panama, Russia, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall number
Z-0793-2019
Recall status
Terminated
CDRH record ID
170550
FDA record created
2019-02-01
Recall event number
81975

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗