Philips Medical Systems North America Co. Phillips
FDA classification date
Dec 21, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
The Accessory Cart contained magnetic material and is attracted to the magnet when bringing it close to the system.
Source facts
Action
On 05/12/2009, Philips began sending out the URGENT-Field Safety Notice MR Accessory Cart to their consignees. The letter identified the affected product and stated that the drawer rails were made from magnetic material and were not according specification. The hazards involved were also provided along with how to identify the affected product.
The firm issued the FCO 78100296 and the firm's Field Service Engineer replaced the magnetic rails with the non-magnetic rails at the consignee's site.
The consignees can contact contact the Philips Healthcare Call Center at 800-722-9377 #5 (then follow prompts) and reference FCO 78100296 with any communication.
Initiated
2009-05-12
Root cause
Nonconforming Material/Component
Terminated
2010-12-28
Distribution
Nationwide Distribution -- IL, MI, and MD.
Recall number
Z-0794-2011
Recall status
Terminated
CDRH record ID
95927
FDA classified
2010-12-21
Recall event number
57351