Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FDA record creation date
Feb 2, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
There is an error in how the programmer calculates and displays the remaining longevity value during a period of time (up to 2 years) prior to the device reaching its Recommended Replacement Time (RRT).
Source facts
Action
In the US, beginning 17-Jan-2019, Medtronic Field Representatives hand delivered a courtesy FCA Notification Letter to consignees who, according to Medtronic records, have programmers that are eligible for Micra software updates. Field Representatives will update the identified programmers using a provided USB. Notification will also be hand delivered to Micra Implanting or Follow-up physicians for awareness of the issue.
Initiated
2019-01-17
Root cause
Software design
Terminated
2021-02-02
Distribution
US, Armenia, Bulgaria, Croatia, Georgia, Hungary, India, Iran, Jordan, Kazakhstan, Latvia, Lebanon, Oman, Pakistan, Qatar, Romania, Saudi Arabia,
Syrian Arab Republic
Recall number
Z-0795-2019
Recall status
Terminated
CDRH record ID
170393
FDA record created
2019-02-02
Recall event number
81949