Open original FDA record ↗Recall record
- FDA identifier
- Z-0797-06
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Apr 26, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Firm became aware of a potential problem that could affect scan results and require patient exam to be repeated when using the customized ECG-gated scan protocols in the CT2006A and CT2006G software of their SOMATOM Sensation CT Systems.
Source facts
- Action
- Firm has issued an advisory letter (3/28/2006) to affected customers per Update Instructions CT008/06/S. This letter informs customers about the potential problem and provides instructions to avoid this issue. A Siemens Service Engineer will visit affected sites to hand-deliver this letter and to delete any affected scan protocols, which can then be replaced with default protocols or with new customized protocols.
- Initiated
- 2006-03-17
- Root cause
- Other
- Terminated
- 2008-04-17
- Distribution
- Nationwide. Product was distributed to medical facilities in AL, AR, AZ, CA, CO, DE, FL, GA, IA, IL IN, KS, KY, MA, MD, MI, MN, NC, NJ, NV, NY, OH, OK, PA, SC, TX, UT and WI.
- Recall number
- Z-0797-06
- Recall status
- Terminated
- CDRH record ID
- 45095
- FDA classified
- 2006-04-26
- Recall event number
- 34995