Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Firm became aware of a potential problem that could affect scan results and require patient exam to be repeated when using the customized ECG-gated scan protocols in the CT2006A and CT2006G software of their SOMATOM Sensation CT Systems.

Source facts

Action
Firm has issued an advisory letter (3/28/2006) to affected customers per Update Instructions CT008/06/S. This letter informs customers about the potential problem and provides instructions to avoid this issue. A Siemens Service Engineer will visit affected sites to hand-deliver this letter and to delete any affected scan protocols, which can then be replaced with default protocols or with new customized protocols.
Initiated
2006-03-17
Root cause
Other
Terminated
2008-04-17
Distribution
Nationwide. Product was distributed to medical facilities in AL, AR, AZ, CA, CO, DE, FL, GA, IA, IL IN, KS, KY, MA, MD, MI, MN, NC, NJ, NV, NY, OH, OK, PA, SC, TX, UT and WI.
Recall number
Z-0797-06
Recall status
Terminated
CDRH record ID
45095
FDA classified
2006-04-26
Recall event number
34995

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗