Z-0797-2016: Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512) Intended for posterior, non-cervical

Recall record Z-0797-2016 for Orthofix, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0797-2016
Company in FDA record
Orthofix, Inc
FDA classification date
Feb 11, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a possibility that the device may deliver a torque value less than the required setting after initial sterilization. In worst case, the unexpected drop in torque value may lead to the spinal construct loosening and requiring corrective surgical intervention.

Source facts

Action
Affected consignees were notified via an Urgent Medical Device Recall Notification letter, dated 1/8/16. The letter identified the affected device and stated the reason for the recall. The recalling firm requests customers to remove the affected devices from their inventory and cease further distribution or use. Customers whom the devices were further distributed to should be notified that the affected devices cannot be used and must be returned to Orthofix per the instructions provided. The attached Acknowledgement Form should be completed and returned. Customers can contact their local Orthofix representative or customer service for further information, replacements, or disposal instructions.
Initiated
2016-01-08
Root cause
Component design/selection
Terminated
2016-11-16
Distribution
Worldwide Distribution -- US, Spain, Italy, and Australia.
Recall number
Z-0797-2016
Recall status
Terminated
CDRH record ID
142988
FDA classified
2016-02-11
Recall event number
73069

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗