Open original FDA record ↗Recall record
- FDA identifier
- Z-0797-2016
- Company in FDA record
- Orthofix, Inc
- FDA classification date
- Feb 11, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a possibility that the device may deliver a torque value less than the required setting after initial sterilization. In worst case, the unexpected drop in torque value may lead to the spinal construct loosening and requiring corrective surgical intervention.
Source facts
- Action
- Affected consignees were notified via an Urgent Medical Device Recall Notification letter, dated 1/8/16. The letter identified the affected device and stated the reason for the recall. The recalling firm requests customers to remove the affected devices from their inventory and cease further distribution or use. Customers whom the devices were further distributed to should be notified that the affected devices cannot be used and must be returned to Orthofix per the instructions provided. The attached Acknowledgement Form should be completed and returned. Customers can contact their local Orthofix representative or customer service for further information, replacements, or disposal instructions.
- Initiated
- 2016-01-08
- Root cause
- Component design/selection
- Terminated
- 2016-11-16
- Distribution
- Worldwide Distribution -- US, Spain, Italy, and Australia.
- Recall number
- Z-0797-2016
- Recall status
- Terminated
- CDRH record ID
- 142988
- FDA classified
- 2016-02-11
- Recall event number
- 73069