Skip to content

Public FDA record

Z-0798-03: Deltec Cozmo Insulin Pump, Model 1700, Product Catalog Numbers 21-1711-01, 21-1712-01 and 21-1713-01. Th

Recall record Z-0798-03 for Deltec, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0798-03
Company in FDA record
Deltec, Inc
FDA classification date
May 9, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A software error could have resulted in the delivery of a larger bolus of insulin than anticipated when the pump was programmed in a particular manner.

Source facts

Action
Recall Notification letters were sent to the consignees on March 14, 2003. The letters gave recommendations for avoiding the problem, stated that updated software to prevent the problem would be introduced, and requested that the customer call a toll-free phone number to arrange replacement of their affected pump.
Initiated
2003-03-14
Root cause
Other
Terminated
2004-07-02
Distribution
The product was distributed nationwide in the United States. There are no foreign or U.S. Government accounts.
Recall number
Z-0798-03
Recall status
Terminated
CDRH record ID
26470
FDA classified
2003-05-09
Recall event number
25780

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗