Skip to content

Public FDA record

Z-0798-2010: 2-Channel Laryngeal Surface Electrode (Model LSE500DCS and LSE500DCL) For use as a Surgical stimulator/lo

Recall record Z-0798-2010 for Neurovision Medical Products Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0798-2010
Company in FDA record
Neurovision Medical Products Inc
FDA classification date
Feb 22, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

This recall has been initiated due to a conductive glue used to attach the lead wire to electrode, which has been found in a small number of electrodes, to cause a short circuit in one of the channels. This is due to migration of the adhesive from one pad to the other creating an electrical short. Use of one of these defective electrodes could lead to a failure to monitor. Our investigation has c

Source facts

Action
URGENT RECALL: Medical Device Letter dated 1/7/10. The letter identified the affected product and provided a list of necessary actions for customers to take. The customers should immediately examine their inventory and quarantine product subject to recall. In addition, if they have further distributed this product, please identify their customers and notify them at once of this product recall. Their notification to their customers may be enhanced by including a copy of this recall notification letter. This recall should be carried out to the user level. Your assistance is appreciated and necessary to prevent any failure to monitor. Please complete and return the enclosed response form as soon as possible. If you have any questions, call Ryan Rea or Kevin Rea, at 866-815-6999.
Initiated
2010-01-11
Root cause
Other
Terminated
2011-04-21
Distribution
Nationwide: CA CO FL GA IL LA MI ND NY OH PA TX UT WA WI)
Recall number
Z-0798-2010
Recall status
Terminated
CDRH record ID
87916
FDA classified
2010-02-22
Recall event number
54233

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗