Open original FDA record ↗Recall record
- FDA identifier
- Z-0798-2020
- Company in FDA record
- CooperSurgical, Inc.
- FDA record creation date
- Jan 10, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for unsealed pouches in two (2) lots of the Endosee Advance Cannula (P/N: ESPX5).
Source facts
- Action
- On December 13, 2019 the firm sent out letters via Fed-Ex stating the following: URGENT: MEDICAL DEVICE RECALL COOPERSURGICAL ENDOSEE CANNULA Dear Valued CooperSurgical Customer, CooperSurgical is issuing a Medical Device Recall for two lot numbers (268748 & 269360) of its ENDOSEE Advance Cannula (CooperSurgical part number: ESPX5) (the Product), because the finished seal (opposite the chevron seal) of the sterile pouch may be compromised, thereby increasing the risk of infection. The Product is an Ethylene Oxide (ETO) sterilized, single-use, disposable cannula, which is sold in 5 Units/box. Each Unit contains an individually wrapped cannula and an individually pouched endometrial suction pipet curette. The compromised seals were detected during a complaint investigation. No adverse events have been reported to CooperSurgical due to this potential issue. This recall affects 2,285 units of the 2 lot numbers shown above, which were manufactured in April and August 2019. We have identified the root cause, and a corrective action has been initiated to prevent future recurrence. Our records indicate that you have purchased the affected Product from CooperSurgical. As indicated in the Instructions-for-Use (IFU), each package should be handled with care and inspected for damage, including the seal area, prior to use. A Product is acceptable for use if it is visually confirmed that the pouchs seal is intact. Please discontinue use of and quarantine Products with any packaging irregularities, complete the attached Acknowledgement and Receipt Form, and return it to CooperSurgical. Once we receive the Acknowledgement Form, we will arrange for Product replacements and returns of affected Product. The replacement Products will be charged at the same price as the original order and your account will be credited back once the affected products are returned to CooperSurgical. If you do not have affected Product in inventory, please use the same enclosed Form to
- Initiated
- 2019-12-13
- Root cause
- Other
- Terminated
- 2021-05-11
- Distribution
- US: AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, WA, WI, WV, & WY. OUS: None
- Recall number
- Z-0798-2020
- Recall status
- Terminated
- CDRH record ID
- 178331
- FDA record created
- 2020-01-10
- Recall event number
- 84443