Skip to content

Public FDA record

Z-0798-2020: ADVANCED ENDOSEE System CANNULA, Cooper Surgical Product Code/Model Number: ESPX5 The Advanced Endosee C

Recall record Z-0798-2020 for CooperSurgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0798-2020
Company in FDA record
CooperSurgical, Inc.
FDA record creation date
Jan 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for unsealed pouches in two (2) lots of the Endosee Advance Cannula (P/N: ESPX5).

Source facts

Action
On December 13, 2019 the firm sent out letters via Fed-Ex stating the following: URGENT: MEDICAL DEVICE RECALL COOPERSURGICAL ENDOSEE CANNULA Dear Valued CooperSurgical Customer, CooperSurgical is issuing a Medical Device Recall for two lot numbers (268748 & 269360) of its ENDOSEE Advance Cannula (CooperSurgical part number: ESPX5) (the Product), because the finished seal (opposite the chevron seal) of the sterile pouch may be compromised, thereby increasing the risk of infection. The Product is an Ethylene Oxide (ETO) sterilized, single-use, disposable cannula, which is sold in 5 Units/box. Each Unit contains an individually wrapped cannula and an individually pouched endometrial suction pipet curette. The compromised seals were detected during a complaint investigation. No adverse events have been reported to CooperSurgical due to this potential issue. This recall affects 2,285 units of the 2 lot numbers shown above, which were manufactured in April and August 2019. We have identified the root cause, and a corrective action has been initiated to prevent future recurrence. Our records indicate that you have purchased the affected Product from CooperSurgical. As indicated in the Instructions-for-Use (IFU), each package should be handled with care and inspected for damage, including the seal area, prior to use. A Product is acceptable for use if it is visually confirmed that the pouchs seal is intact. Please discontinue use of and quarantine Products with any packaging irregularities, complete the attached Acknowledgement and Receipt Form, and return it to CooperSurgical. Once we receive the Acknowledgement Form, we will arrange for Product replacements and returns of affected Product. The replacement Products will be charged at the same price as the original order and your account will be credited back once the affected products are returned to CooperSurgical. If you do not have affected Product in inventory, please use the same enclosed Form to
Initiated
2019-12-13
Root cause
Other
Terminated
2021-05-11
Distribution
US: AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, WA, WI, WV, & WY. OUS: None
Recall number
Z-0798-2020
Recall status
Terminated
CDRH record ID
178331
FDA record created
2020-01-10
Recall event number
84443

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗