Open original FDA record ↗Recall record
- FDA identifier
- Z-0799-2013
- Company in FDA record
- Fresenius Medical Care Holdings, Inc.
- FDA classification date
- Feb 11, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a small risk of a leak in Cycler Sets and if the Cycler Set is punctured, the dialysate may become contaminated potentially resulting in peritonitis.
Source facts
- Action
- Fresenius Medical Care began contacting the affected customers by telephone and mailing an "URGENT: FMCNA LIBERTY CYCLER SET RECALL" letter and Return Receipt form dated October 31, 2012. The letters identified the specific part number and product lot and request that customers discontinue use of the lots immediately and to return unused product to FMCNA. Instructions are provided for returning the product, asking customers to isolate and discontinue use of the lot number 12BR08009 and place the product in a secure, segregated area. Consignees were instructed to contact their Fresenius Medical Care Customer Service Team at 1-800-323-5188 for instructions on how to return the recalled product.
- Initiated
- 2012-10-12
- Root cause
- Process control
- Terminated
- 2020-12-30
- Distribution
- Nationwide Distribution.
- Recall number
- Z-0799-2013
- Recall status
- Terminated
- CDRH record ID
- 115494
- FDA classified
- 2013-02-11
- Recall event number
- 64175