Open original FDA record ↗Recall record
- FDA identifier
- Z-0799-2014
- Company in FDA record
- Zeiss, Carl Inc
- FDA classification date
- Jan 21, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Carl Zeiss Microscopy, LLC has recalled the Compensator Slider Lamda 6x20 component utilized on Standard Pathology Microscopes due to misalignment of the lambda plate.
Source facts
- Action
- Carl Zeiss, Inc. sent an Important Medical Device Corrective Action letter dated December 20, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised that their service provider would contact them to schedule a visit in which a factory trained service technician would install a new Compensator Slider Lambida 6x20. Customers should return the old Compensator Slider to the factory for testing. Customers with technical questions were instructed to call 914-681-7815.
- Initiated
- 2013-12-20
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-04-08
- Distribution
- Worldwide Distribution - Nationwide including NY, TX, NC, FL, ME, MA and MI; and Internationally to Arab Emirates, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, China, Czech Republic, Egypt, France, Germany, Hong Kong, India, Italy, Kosovo, Lithuania, Mexico, The Netherlands, New Zealand, Peru, Poland, Spain, and Taiwan.
- Recall number
- Z-0799-2014
- Recall status
- Terminated
- CDRH record ID
- 124877
- FDA classified
- 2014-01-21
- Recall event number
- 67178