Open original FDA record ↗Recall record
- FDA identifier
- Z-0799-2016
- Company in FDA record
- Cook Inc.
- FDA classification date
- Feb 24, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
A manufacturing process may lead to catheter tip fracture and or separation. Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding blood flow to end organs. These catheters are used in venous or arterial pressure monitoring, blood sampling, and administration of drugs and fluids.
Source facts
- Action
- On 1/6/2016, URGENT: MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital Risk Management/Recall Administration via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Should you have any questions or concerns, please do not hesitate to contact Cook Medical Customer Relations at (800) 457-4500 or 1(812) 339-2235.
- Initiated
- 2016-01-06
- Root cause
- Process control
- Terminated
- 2017-08-23
- Distribution
- Worldwide Distribution - US Nationwide including DC and PR including the countries of Australia, Germany, China, Chile, Dominican Republic, Panama, Israel, Brazil, Canada
- Recall number
- Z-0799-2016
- Recall status
- Terminated
- CDRH record ID
- 142930
- FDA classified
- 2016-02-24
- Recall event number
- 73056