Skip to content

Public FDA record

Z-0799-2020: Gridlock Ankle Screw Driver Bit REF 320-35-003 - Product Usage: A screwdriver facilitates implant inserti

Recall record Z-0799-2020 for Trilliant Surgical, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0799-2020
Company in FDA record
Trilliant Surgical, LLC
FDA record creation date
Jan 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A screwdriver used to facilitate implant insertion and removal was found to be out of specification for the minor diameter of the drive feature causing the driver to be unable to retain screws as intended.

Source facts

Action
The recall strategy included sending written recall communication and response forms as of 03/14/2019 via FedEx (Express Standard Overnight) to all consignees that showed possession / had been sold a device (as the device is considered reusable) as of 03/14/2019. The communication instructed the consignees to: " Check inventory stock for identified devices " Locate and remove and/or discontinue use of identified devices " Return all related product(s) to Recalling Firm in accordance with the included return response / acknowledgement form.
Initiated
2019-03-14
Root cause
Process control
Terminated
2022-03-16
Distribution
US Nationwide distribution in the states of CO, CT, IA, TX and WI. OUS: None
Recall number
Z-0799-2020
Recall status
Terminated
CDRH record ID
178110
FDA record created
2020-01-10
Recall event number
84350

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗