Open original FDA record ↗Recall record
- FDA identifier
- Z-0799-2020
- Company in FDA record
- Trilliant Surgical, LLC
- FDA record creation date
- Jan 10, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A screwdriver used to facilitate implant insertion and removal was found to be out of specification for the minor diameter of the drive feature causing the driver to be unable to retain screws as intended.
Source facts
- Action
- The recall strategy included sending written recall communication and response forms as of 03/14/2019 via FedEx (Express Standard Overnight) to all consignees that showed possession / had been sold a device (as the device is considered reusable) as of 03/14/2019. The communication instructed the consignees to: " Check inventory stock for identified devices " Locate and remove and/or discontinue use of identified devices " Return all related product(s) to Recalling Firm in accordance with the included return response / acknowledgement form.
- Initiated
- 2019-03-14
- Root cause
- Process control
- Terminated
- 2022-03-16
- Distribution
- US Nationwide distribution in the states of CO, CT, IA, TX and WI. OUS: None
- Recall number
- Z-0799-2020
- Recall status
- Terminated
- CDRH record ID
- 178110
- FDA record created
- 2020-01-10
- Recall event number
- 84350