Z-0800-05: Baxter Spinal Anesthesia Trays; sterile, single use only procedure trays that contain BD 25G x 8.9 cm (3-

Recall record Z-0800-05 for Baxter Healthcare Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0800-05
Company in FDA record
Baxter Healthcare Corp.
FDA classification date
May 6, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The anesthesia trays contain BD 25G Spinal Needles which which have leakage at the plastic hub of the needle.

Source facts

Action
Urgent Device Correction letters dated 3/28/05 were sent to the direct accounts via first class mail. The accounts were informed of the leakage associated with the plastic hub of the spinal needle contained in the trays, and were provided with a listing of product codes and lots numbers of the spinal trays affected. The accounts were instructed to check their inventory for the affected trays and to contact BD at 1-888-237-2762 for instructions on how to obtain replacement spinal needles. Replacement needles will be shipped overnight, and the accounts were instructed to remove and destroy the needle in the tray and replace it with the replacement needles when the tray is opened prior to the procedure.
Initiated
2005-03-28
Root cause
Other
Terminated
2008-03-04
Distribution
Nationwide, Puerto Rico, Amaerical Samoa, Guam, Canada and Venezuela
Recall number
Z-0800-05
Recall status
Terminated
CDRH record ID
37982
FDA classified
2005-05-06
Recall event number
31491

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗