Open original FDA record ↗Recall record
- FDA identifier
- Z-0800-2014
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- Jan 21, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An overflow flag //// is displayed in the result column for a test when the algorithm for the test cannot calculate a result or when the concentration cannot be determined because it is outside the absorbance limits, and it is possible for the user to misinterpret or edit the result.
Source facts
- Action
- The distribution of the Urgent Medical Device Correction letter scheduled to begin on December 17, 2013. x Field service personnel will be notified of the Urgent Medical Device Correction Letter via the Siemens Knowledge Base, as well as information regarding placing labels on the customer system to remind the customer that ///// is not a reportable result. x The Customers were instructed to contact Siemens Healthcare Diagnostics Technical Solution Center if they have technical questions regarding this issue. x Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction.
- Initiated
- 2013-12-17
- Root cause
- Software design
- Terminated
- 2014-08-26
- Distribution
- Worldwide Distribution. USA nationwide and the country of Canada.
- Recall number
- Z-0800-2014
- Recall status
- Terminated
- CDRH record ID
- 124573
- FDA classified
- 2014-01-21
- Recall event number
- 67030