Open original FDA record ↗Recall record
- FDA identifier
- Z-0800-2015
- Company in FDA record
- Carestream Health Inc
- FDA classification date
- Dec 16, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
While servicing a unit at a customer site, a field engineer discovered a broken bolt from the carriage assembly. Carestream has identified an issue related to the DRX-Revolution Mobile X-Ray System in which bolts that secure the x-ray tube assembly/boom to the device mounting block may break. It has been determined that this issue is a result of an improperly assembled part which holds the x-ray
Source facts
- Action
- The firm, Carestream, sent an "URGENT MEDICAL DEVICE RECALL" letter dated 11/20/2014 to US Consignees/customers via Certified Mail on 11/25/2014. Foreign accounts will be notified as soon as possible. The letter described the product, problem and actions to taken. The customers were instructed if they observe a loosened boom to follow the procedure for turning the system off; stop using the device and contact Carestream Customer Care Center in the U.S. at 1-800-328-2910, available 7 days per week on a 24 hour basis; Outside of the U.S., call your local Service support number. Carestream Health Service representative will contact all customers to schedule a visit to inspect the DRX-Revolution Mobile X-Ray System. If a repair is required it will be completed by a Carestream Field Engineer. If you have any questions, contact the Regulatory Affairs Manager at 585-627-8230 or Carestream Customer Care Center in the U.S. at 1-800-328-2910, available 7 days per week on a 24 hour basis.
- Initiated
- 2014-11-25
- Root cause
- Device Design
- Terminated
- 2016-06-06
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: AUSTRALIA, BELGIUM, CANADA, CHILE, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, ITALY, JAPAN, KOREA, KUWAIT, MALAYSIA, NETHERLANDS, NEW ZEALAND, NORWAY, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, and UNITED KINGDOM.
- Recall number
- Z-0800-2015
- Recall status
- Terminated
- CDRH record ID
- 131774
- FDA classified
- 2014-12-16
- Recall event number
- 69826