Z-0800-2020: QuickDraw Venous Cannula, REF: QD22, UDI: 00690103182699; and REF: QD25, UDI: 00690103182705

Recall record Z-0800-2020 for Edwards Lifesciences, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0800-2020
Company in FDA record
Edwards Lifesciences, LLC
FDA record creation date
Jan 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If the venous cannula is used in an unintended manner (longer than 6 hours; particularly during extracorporeal membrane oxygenation [ECMO]) there is a risk that separation of the cannula from its connector can occur; which can result in significant blood loss.

Source facts

Action
On 11/27/19, Urgent Medical Device Safety Notices were mailed to customers (Attention: Risk Management Departments). Customers were informed the following: 1. Review the field safety notice to understand the potential hazard. 2. Meet and review with the appropriate clinical staff at your hospital. 3. Complete and return the acknowledgment form attached to this letter via fax (949-250-2525). 4. Distribute this notice within your organization or to any organization where the devices under the safety alert have been transferred. 5. Make the information available to personnel that may be using these devices as part of continuing communication and training. 6. Instructions for Use are being modified to add a contraindication for long term use (greater-than 6 hours, particularly during extracorporeal membrane oxygenation (ECMO) procedures. 7. Product return is not required. If you have additional questions, please contact customer service at 800-424-3278, Option 1, from 5:00 AM - 4:30 PM PST.
Initiated
2019-11-27
Root cause
Other
Terminated
2024-05-28
Distribution
U.S.: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV. O.U.S. (Foreign): Brazil, Colombia, Albania, Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Czech Republic, Denmark, France, Germany, Greece, Ireland, Israel, Italy, Japan, Jordan, Malaysia, Namibia, Poland, Romania, Portugal, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Recall number
Z-0800-2020
Recall status
Terminated
CDRH record ID
178198
FDA record created
2020-01-10
Recall event number
84381

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗