A change in the automatic capacitor reform schedule during the later portion of the device's useful life can result in excessive 750V charge times and disable the device's ability to report an End of Life (EOL) indicator due to excessive charge times unless the patient is receiving periodic 750V therapy shocks.
Source facts
Action
Angeion notified physicians responsible for managing each remaining implanted ICD, also provided them with programming means to change the automatic cap charging schedule, and to monitor and document the reprogramming process.
Initiated
2005-01-25
Root cause
Other
Terminated
2006-07-22
Distribution
AZ, CA, GA, IL, IN, KY, MA, MD, MI, MN, NC, NJ, OH, PA and to UK, Italy, Germany and Argentina.
Recall number
Z-0801-05
Recall status
Terminated
CDRH record ID
38111
FDA classified
2005-05-06
Recall event number
31539