Z-0801-2020: BriteBlade Pro Single-Use Fiber Optic Mac 3, CE01120, Do Not Re-use, Rx Only, REF 040-713U, Sterile EO, Q

Recall record Z-0801-2020 for Flexicare Medical Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0801-2020
Company in FDA record
Flexicare Medical Ltd.
FDA classification date
Jan 13, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.

Source facts

Action
On 11/27/2019, the firm "Flexicare" sent a "Urgent: medical Device Recall" letter via USPS Priority mail. The letter informed consignees of the recall. In addition to informing consignees of the recall the letter asked consignees to take the following actions: Complete the acknowledgment and receipt form and return this form to Flexicare Inc. On receipt of the form you will be contacted by a representative of Flexicare Inc. to make all necessary further arrangements. The form asked the following of the consignee: -Acknowledge that you have read and understand the recall instructions provided - If you have experienced any adverse events associated with the recalled products. - Check your stock and quarantined inventory - Send completed response forms marked for the attention of Product Manager Anesthesia to email Ian.ross@flexicare.com or by fax to 949-450-9992 - Identify and notify your customers that were shipped product from the affected lot numbers If you have any questions, please call Product Manager Anesthesia, at 949-450-9999 ext. 102 or email at Ian.Ross@flexicare.com.
Initiated
2019-11-27
Root cause
Nonconforming Material/Component
Terminated
2021-04-27
Distribution
US: Nationwide OUS: Australia, Canada, China, Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK
Recall number
Z-0801-2020
Recall status
Terminated
CDRH record ID
178156
FDA classified
2020-01-13
Recall event number
84367

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗