Open original FDA record ↗Recall record
- FDA identifier
- Z-0801-2021
- Company in FDA record
- Ortho-Clinical Diagnostics, Inc.
- FDA classification date
- Jan 7, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A software anomaly may cause an aliquoted sample to be labelled as the incorrect sample. This may lead to results being associated with the wrong patient sample and potentially lead to inappropriate intervention with the potential for injury to the patient.
Source facts
- Action
- On November 20, 2020, the firm sent Urgent Product Correction Notification letters to affected customers to inform them of the issue. Customers were reminded to follow product labelling regarding removing all tubes before performing maintenance or initializing the Aliquoter & Labeller Module. There should never be sample tubes present when initializing the Aliquoter & Labeller module. The firm's release of TCA Software version 4.7 is the planned mitigation for this issue. If you have questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
- Initiated
- 2020-11-20
- Root cause
- Software design
- Terminated
- 2023-10-24
- Distribution
- Worldwide distribution - US Nationwide distribution including the states of CA, CO, MA, MD, NC, OH, TX, VA, VT, and Belgium, Bermuda, Canada, China, France, Italy, Japan, Mexico, Norway, Portugal, Singapore, Spain, Sweden, and United Kingdom.
- Recall number
- Z-0801-2021
- Recall status
- Terminated
- CDRH record ID
- 184956
- FDA classified
- 2021-01-07
- Recall event number
- 86972