Potential for internal blood leaks due to cracked polyurethane
Source facts
Action
Fresenius Medical Care notified consignees on about 12/05/2024 via URGENT MEDICAL DEVICE RECALL letter. Consignees were instructed to stock immediately for either of the affected lot numbers, immediately discontinue use of Optiflux F160NRe Dialyzer(s) with lot codes, place any of the affected units in a secure, segregated area, arrange for return of affected product and complete and return the provided Reply Form. Additionally, consignees were instructed to forward the notification to any personnel within the organization or to customers where product was further distributed.
Initiated
2024-12-05
Root cause
Process control
Distribution
Nationwide Distribution.
Recall number
Z-0801-2025
Recall status
Open, Classified
CDRH record ID
211440
FDA classified
2025-01-03
Recall event number
95884