Inaccurate readings. Software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations. The accuracy on any cuvette size did not meet the specifications.
Source facts
Action
Terumo sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated November 1, 2010 to all affected customers. The letter identified the product, problem, and the actions to be taken by the customers. Contact Terumo CVS Customer Service at 1-800-521-2818 for questions or concerns.
Initiated
2010-11-01
Root cause
Software design
Terminated
2013-11-08
Distribution
Worldwide Distribution.
Recall number
Z-0802-2013
Recall status
Terminated
CDRH record ID
106244
FDA classified
2013-02-12
Recall event number
60695