The firm discovered that during the use of these defibrillation electrodes a possibility for arcing and a resulting malfunction exists. This could lead to a situation, in which a patient who is in a life threatening situation requiring a defibrillation can receive such therapy only with a delay or not at all.
Source facts
Action
The one consignee was sent an Urgent Medical Device Recall letter, dated 12/3/2014. The letter identified affected product, explained the issue, and asked for product to stop being used and distributed, and to return the devices. Questions can be addressed to Bettina Sarlay, Sales Manager, Leonhard Lang, at bettina.sarlay@leonhardlang.at.
Initiated
2014-12-03
Root cause
Component design/selection
Terminated
2015-08-13
Distribution
US Nationwide Distribution in the states of OH and TX.
Recall number
Z-0802-2015
Recall status
Terminated
CDRH record ID
131992
FDA classified
2014-12-19
Recall event number
69930