Inability or difficulty in unbuckling/releasing the proning arm buckle might delay patient access in an emergency situation such as cardiac arrest. This, in turn, could contribute to a life-threatening event.
Source facts
Action
Arjo hand-delivered the notice with each RotoProne Therapy System rental and delivery beginning on 12/19/2018. Arjo will require the customer to confirm the notification has been received and understood via the signed delivery notification.
Initiated
2018-12-19
Root cause
Component design/selection
Terminated
2022-01-26
Distribution
US, Saudi Arabia, and Kuwait
Recall number
Z-0802-2019
Recall status
Terminated
CDRH record ID
170171
FDA classified
2019-02-05
Recall event number
81872
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.