Skip to content

Public FDA record

Z-0802-2021: Pinnacle Cups, Part numbers 121701048 121701050 121701052 121701054 121701056 121701058 121701060 1217010

Recall record Z-0802-2021 for DePuy Orthopaedics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0802-2021
Company in FDA record
DePuy Orthopaedics, Inc.
FDA record creation date
Jan 7, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain Pinnacle Cup devices may potentially exhibit an oversized "minor diameter", which could lead the Apex HE to thread through the shell of the cup without stopping or to protrude internally as a result of "cross-threading". If cross-threading occurs, physicians will not feel the positive stop of the Apex HE against the cup.

Source facts

Action
On December 15, 2020, the firm distributed Urgent Medical Device Correction letters to customers. The letter informed customers that there is a potential for an oversized "minor diameter" on certain Pinnacle Cup devices. The correction is intended to advise physicians of the oversize condition of the hole in the cup, which could lead the Apex HE to thread through the shell of the cup without stopping or to protrude internally as a result of cross-threading. This may occur if the Apex HE is manipulated or toggled while threading it through the hole of certain Pinnacle Cup devices. If cross-threading occurs, physicians will not feel the positive stop of the Apex HE against the cup. Customers were asked to review the information in the letter and complete and return the included customer response form. If any of the affected products have been forwarded to another facility, please contact that facility and provide a copy of the recall notification to the relevant personnel. Post a copy of the notice in a visible area for awareness. If you have any questions, please contact your Sales Consultant.
Initiated
2020-12-15
Root cause
Process control
Terminated
2023-11-14
Distribution
Worldwide distribution - US Nationwide distribution.
Recall number
Z-0802-2021
Recall status
Terminated
CDRH record ID
184895
FDA record created
2021-01-07
Recall event number
86946

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗