Stryker has received a complaint from customers relating to a fracture during the use of the Specialty Audible Torque Wrench in surgery.
Source facts
Action
Stryker Spine sent an Urgent Medical Device Removal Notification/Acknowledgement Form dated 10/6/2014 via Fed Ex. The letter identified the affected product, discussed the product description, product issue, potential hazards, risk mitigations, and actions needed. Customers were asked to read the notice thoroughly and make sure that it is understood. Additionally, they are to complete the enclosed acknowledgement form and fax or e-mail it back.
Initiated
2014-10-06
Root cause
Nonconforming Material/Component
Terminated
2017-05-08
Distribution
Distributed to GA, IN, MI, NJ, OH, PA, TX & VA.
Recall number
Z-0803-2015
Recall status
Terminated
CDRH record ID
131536
FDA classified
2014-12-16
Recall event number
69784