- Recall classification
- Class III
Reason stated in the source
Some material of the coating Type CXCPMV3 has potential artifacts showing a small bright thin line when the film is processed.
Source facts
- Action
- Dealers were contacted by Agfa HealthCare via phone and/or email on 10/24/2014 and the week of 11/10/14 to inform them of the quality issue and to quarantine identified product shipped to their facility or end users. On 11/17/2014, an "URGENT SAFETY NOTICE" letter was sent to the US dealers. The letter described the safety alert and the request to quarantine affected product. Acknowledgment via FAX-Back or email that the information was received and understood was requested from the dealers. A customer Urgent Field Safety Notice was also provided to those dealers who had shipped affected product to end users so they could sent it to the end use customers.
- Initiated
- 2014-10-24
- Root cause
- Other
- Terminated
- 2015-03-16
- Distribution
- US Nationwide Distribution in the states of: AL, AZ, CA, LA, MI, NE, NH, NJ, NC, VA
- Recall number
- Z-0804-2015
- Recall status
- Terminated
- CDRH record ID
- 131721
- FDA classified
- 2014-12-16
- Recall event number
- 69864