Z-0804-2018: STAT 2 I.V. Gravity Flow Controller and STAT 2 Primary, Secondary Administration Sets and Extension Sets

Recall record Z-0804-2018 for ConMed Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0804-2018
Company in FDA record
ConMed Corporation
FDA record creation date
Feb 28, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Intravascular (I.V.) fluid administration sets are exhibiting inaccurate or inconsistent flow rates.

Source facts

Action
The firm, ConMed, sent an "URGENT: DEVICE RECALL" letter dated 6/21/2017 to its domestic and foreign customers. The letter described the product, the problem and the actions to be taken. The customers were instructed NOT to use any STAT 2 I.V. Gravity Flow Controllers manufactured between April 27, 2012 and April 12, 2017; review your inventory for any of the devices listed on Attachment I; contact all of those departments within your facility and any other facilities within your organization that may have received affected products. It is imperative that all end users of these devices receive this notice and respond immediately; if you HAVE inventory of any of the devices listed, complete the Business Reply Form and return it with devices to: CONMED Corporation, 525 French Road, Utica, NY 13502 via UPS Account # W5Y243; if you DO NOT HAVE any affected devices complete and return the form by email: stat2@conmed.com; fax: Field Action Support Team at 315-624-3225 or US mail to CONMED Corporation, 525 French Road, Utica, NY 13502;Attn: Field Action Support Team. PLEASE DO NOT RETURN USED DEVICES. If you have any questions or requests, please dont hesitate to contact the Field Action Support Team at 1-800-448-6506, fax to 315-624-3225, or email stat2@conmed.com.
Initiated
2017-06-19
Root cause
Nonconforming Material/Component
Terminated
2022-07-25
Distribution
Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, Washington, DC, Puerto Rico, and countries of: Lebanon, Panama, Saudi Arabia, Germany, India, and Canada.
Recall number
Z-0804-2018
Recall status
Terminated
CDRH record ID
161416
FDA record created
2018-02-28
Recall event number
79129

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗