Z-0804-2019: Richard Wolf Bronchoscope Tubes, Model Texas. Device is reusable and should be sterilized prior to each

Recall record Z-0804-2019 for Richard Wolf GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0804-2019
Company in FDA record
Richard Wolf GmbH
FDA record creation date
Feb 6, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential deficiency of the TEXAS Bronchoscope Tube product. The non-functionally relevant bonding points / cemented areas of the Bronchoscope Tubes may loosen.

Source facts

Action
On about 09/24/2018, Richard Wolf GmbH sent Urgent Safety Information notices to international consignees. Customers in the USA were sent the "Urgent Safety Information" notice on about 10/17/2018. Customers were instructed to check current stock on hand for affected Bronchoscope Tubes, complete and return the attached response form, ensure all appropriate personnel at the facility are notified of the recall, ensure the recall information is observed in your respective institution until all internal measures have been completed, notify Richard Wolf of any adverse events, and return all affected devices included in the recall. Richard Wolf GmbH expanded the recall and notified customers worldwide on about 12/06/2018 via "Urgent Safety Information" notice. Instructions included to check current stock on hand for affected product and lot number, return affected products to Richard Wolf GmbH, notify customers if affected product was further distributed, ensure the recall notification is observed in each customer location, and complete and return the Answer Form provided. Richard Wolf Medical Instruments (US Distributor) notified customers on about 12/14/2018 via "Urgent Safety Information: Recall Expansion" notice. Instructions included to check current stock on hand for affected product and lot number, return affected products to Richard Wolf Medical Instruments, notify customers if affected product was further distributed, ensure the recall notification is observed in each customer location, and complete and return the Answer Form provided.
Initiated
2018-10-17
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide in the states of PA, MD, NY, and countries of Bangladesh, Canada, China, France, Germany, India, Italy, Jordan, Netherlands, Poland, Russia, Spain, Thailand, Turkey.
Recall number
Z-0804-2019
Recall status
Open, Classified
CDRH record ID
167657
FDA record created
2019-02-06
Recall event number
81169

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗