- Class II
Reason stated in the source
The Smaxel Fractional CO2 Laser does not include a Certification label, and adequate instructions for safe operation of the device.
Source facts
- Action
- IDS LTD recommend that users are advised to use it as close to the main body as possible during the procedure. The firm will without charge remedy the defect or bring the product into compliance with each applicable Federal standard. On or about 1/22/2025, the firm mailed a "Notification Letter" to affected customers/users informing them that the SAXEL Fractional CO2 Laser System does not have affixed a certification label and that the firm will be implementing a design change by revising the certification label and will provide the revised certification label free of charge following the design change.
- Initiated
- 2024-10-16
- Root cause
- Radiation Control for Health and Safety Act
- Distribution
- Worldwide - US Nationwide Distribution
- Recall number
- Z-0804-2025
- Recall status
- Open, Classified
- CDRH record ID
- 211821
- FDA classified
- 2025-01-07
- Recall event number
- 96034