Skip to content

Public FDA record

Z-0804-2025: Smaxel Fractional CO2 Laser, a Class IV medical laser system.

Recall record Z-0804-2025 for IDS LTD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0804-2025
Company in FDA record
IDS LTD
FDA classification date
Jan 7, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Smaxel Fractional CO2 Laser does not include a Certification label, and adequate instructions for safe operation of the device.

Source facts

Action
IDS LTD recommend that users are advised to use it as close to the main body as possible during the procedure. The firm will without charge remedy the defect or bring the product into compliance with each applicable Federal standard. On or about 1/22/2025, the firm mailed a "Notification Letter" to affected customers/users informing them that the SAXEL Fractional CO2 Laser System does not have affixed a certification label and that the firm will be implementing a design change by revising the certification label and will provide the revised certification label free of charge following the design change.
Initiated
2024-10-16
Root cause
Radiation Control for Health and Safety Act
Distribution
Worldwide - US Nationwide Distribution
Recall number
Z-0804-2025
Recall status
Open, Classified
CDRH record ID
211821
FDA classified
2025-01-07
Recall event number
96034

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗