Z-0805-2012: GEMINI TF 16 PET/CT System (with software version 3.5.1 and 3.5.1.1), Model # 882470 & GEMINI TF 64 PET/C

Recall record Z-0805-2012 for Philips Medical Systems (Cleveland) Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0805-2012
Company in FDA record
Philips Medical Systems (Cleveland) Inc
FDA classification date
Jan 18, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Philips received notification by a customer of an issue relating to unwanted auto delete occurrences resulting in raw data being deleted and, in some cases, raw data being deleted immediately after scan acquisition but before images are reconstructed.

Source facts

Action
Philips Medical System sent a Urgent Medical Device Correction letter dated December 5, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The following are recommended actions: 1. Please ask your local Philips Service Engineer to ensure that the Auto Delete function is configured to run automatically. 2. Please share this information with all Philips GEMINI TF system operators at your site. 3. Retain this document with your system's instructions for Use Manual. For further information or support concerning this issue please contact your local Philips representative in the United States and Canada at 1-800-722-9377, and follow the recorded menu options to reach Customer Solutions Engineer; in all other countries please dial your local Philips Healthcare office.
Initiated
2011-12-02
Root cause
Software design
Terminated
2013-02-22
Distribution
Worldwide Distribution -- USA (nationwide) including the states: AL, AR, AZ, CA, CO, DC, FL, GA, HI, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NY, OH, OR, PA, PR, TN, TX, WA and WV. and the countries of: Argentina, Australia, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Denmark, Egypt, France, Germany, Hungary, India, Israel, Italy, Japan, Netherlands, Pakistan, Philippines, Poland, Russian Fed, Saudi, Arabia, Singapore, South Korea, Spain, Sweden, Taiwan, Thailand, Turkey, Ukraine and United Kingdom.
Recall number
Z-0805-2012
Recall status
Terminated
CDRH record ID
106832
FDA classified
2012-01-18
Recall event number
60933

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗